DATASHEET
| E1 | – |
| E2 | – |
| F1 | – |
| F2 | – |
| M1 | 1000 kg |
| M2 | 200 kg — 1000 kg |
| Maximum capacity [Max] | 1050 kg |
| Readability [d] | 10 g |
| Standard repeatability [5% Max] | 6 g |
| Standard repeatability [Max] | 10 |
| Permissible repeatability | 15 |
| Eccentricity (tested load) | 1d / 5 mm |
| Electric compensation range | 0 — 1050 kg |
| Stabilization time | 10 s |
| Adjustment | internal (automatic) |
| Display | 10″ touchscreen |
| Weighing pan dimensions | 1000×800 mm |
| Controling device dimensions | 206×140×70 mm |
| Comparator packaging dimensions without self-centering weighing pan | 1200×1000×440 mm |
| Net weight | 121 kg |
| Gross weight | 161 kg |
| Recommended load speed | 0.3 m/min |
| Maximum permissible load speed | 0.6 m/min |
| Weighing unit dimensions without self-centering weighing pan | 1010×800×182 mm |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Operating temperature | +15 — +30 °C |
| Operating temperature change rate | ±2°C/12h |
| Relative humidity | 40% — 60% |
| Relative humidity change rate | ±10%/4h |
Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of F2 class mass standards specified in OIML R111 (Table C.1.) document.