DATASHEET
| E1 | – |
| E2 | – |
| F1 | 10 kg — 20 kg |
| F2 | 2 kg — 20 kg |
| M1 | 1 kg — 25 kg |
| M2 | 200 g — 25 kg |
| Maximum capacity [Max] | 25.5 kg |
| Readability [d] | 10 mg |
| Standard repeatability [5% Max] | 4 mg |
| Standard repeatability [Max] | 10 mg |
| Permissible repeatability | 15 mg |
| Eccentricity (tested load) | 1d / 2 mm |
| Electric compensation range | 0 — 25.5 kg |
| Stabilization time | 5 s |
| Adjustment | internal |
| Display | 10″ touchscreen |
| Weighing pan dimensions | 302×252 mm |
| Weighing device dimensions | 370×280×150 mm |
| Controlling device dimensions | 206×140×70 mm |
| Packaging dimensions | 520×520×280 mm |
| Net weight | 11.5 kg |
| Gross weight | 13.7 kg |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Operating temperature | +10 — +40 °C |
| Operating temperature change rate | ±2°C/12h |
| Relative humidity | 30% — 70% |
| Relative humidity change rate | ±10%/4h |
Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of F2 class mass standards specified in OIML R111 (Table C.1.) document.