DATASHEET
| E1 | 1 mg — 5 g |
| E2 | 1 mg — 5 g |
| F1 | 1 mg — 5 g |
| F2 | 1 mg — 5 g |
| Maximum capacity [Max] | 6.1 g |
| Readability [d] | 0.0001 mg |
| Standard repeatability [5% Max] | 0.2 µg |
| Standard repeatability [Max] | 0.4 µg |
| Permissible repeatability | 0.6 µg |
| Eccentricity (tested load) | 0 µg |
| Electric compensation range | 0 — 6.1 g |
| Stabilization time | 30 s |
| Adjustment | internal (automatic) |
| Display | 10″ touchscreen |
| Magazine positions | 36 |
| Weighing pan dimensions | ø20 mm |
| Weighing device dimensions | 950 ×590×540 mm |
| Controlling device dimensions | 460×250×195 mm |
| Packaging dimensions | 1200×800×950 mm |
| Net weight | 102.5 kg |
| Gross weight | 183.5 kg |
| Supplementary weights internal | – |
| Supplementary weights external | – |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Power supply | 100-240 V AC 50/60 Hz |
| Operating temperature | +15 — +30°C |
| Operating temperature change rate | ±0.5°C/12h (±0.3°C/4h) |
| Relative humidity | 40% — 60% |
| Relative humidity change rate | ±5%/12h (3%/4h) |
Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of E1 class mass standards specified in OIML R111 (Table C.1.) document.