DATASHEET
| E1 | 500 g — 1 kg |
| E2 | 100 g — 1 kg |
| F1 | 10 g — 1 kg |
| F2 | 1 g — 1 kg |
| M1 | 1 g — 1 kg |
| M2 | 1 g — 1 kg |
| Maximum capacity [Max] | 1.02 kg |
| Readability [d] | 0.01 mg |
| Standard repeatability [5% Max] | 0.25 mg |
| Standard repeatability [Max] | 0.03 mg |
| Permissible repeatability | 0.05 mg |
| Eccentricity (tested load) | 1d / 1 mm |
| Electric compensation range | -10 g — +20 g |
| Stabilization time | 30 s |
| Adjustment | external |
| Display | 10″ touchscreen |
| Weighing pan dimensions | ø60 mm |
| Weighing device dimensions | 385×215×525 mm |
| Controlling device dimensions | 206×140×70 mm |
| Anti-draft chamber dimensions | 660×470×700 mm |
| Packaging dimensions | 860×800×550 mm |
| Chamber packaging size | 820×850×630 mm |
| Net weight | 38 kg |
| Gross weight | 58 kg |
| Supplementary weights internal | half automatic |
| Supplementary weights external | 50 g, 30 g, 2× 10 g |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Operating temperature | +15 — +30 °C |
| Operating temperature change rate | ±0.5°C/12h (±0.3°C/4h) |
| Relative humidity | 40% — 60% |
| Relative humidity change rate | ± 3 % / 4 h |
Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of F2 class mass standards specified in OIML R111 (Table C.1.) document.
Eccentricity (tested load) – suspended self-centring weighing pan (available as an option): 0 mg
* Eccentricity (tested load): 1 d / mm












Ulasan