DATASHEET
| E1 | 1 mg — 5 g |
| E2 | 1 mg — 5 g |
| F1 | 1 mg — 5 g |
| F2 | 1 mg — 5 g |
| M1 | 1 mg — 5 g |
| M2 | 1 mg — 5 g |
| Maximum capacity [Max] | 6.1 g |
| Readability [d] | 0.1 µg |
| Standard repeatability [5% Max] | 0.2 µg |
| Standard repeatability [Max] | 0.4 µg |
| Permissible repeatability | 0.6 µg |
| Eccentricity (tested load) | 1d / 1 mm |
| Electric compensation range | 0 — 6.1 g |
| Stabilization time | 30 s |
| Adjustment | internal (automatic) |
| Display | 10″ touchscreen |
| Weighing pan dimensions | ø16 mm |
| Weighing device dimensions | 370×160×175 mm |
| Controlling device dimensions | 206×140×70 mm |
| Anti-draft chamber dimensions | 560×350×255 mm |
| Packaging dimensions | 655×755×445 mm |
| Chamber packaging size | 820×840×630 mm |
| Net weight | 10.1 kg |
| Gross weight | 17.6 kg |
| Supplementary weights internal | – |
| Supplementary weights external | – |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Operating temperature | +15 — +30°C |
| Operating temperature change rate | ±0.5°C/12h (±0.3°C/4h) |
| Relative humidity | 40% — 60% |
| Relative humidity change rate | ± 2 % / 4 h |
Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of E1 class mass standards specified in OIML R111 (Table C.1.) document.