ALIBI Memory
Data safety is ensured by the ALIBI memory integrated into AS R1 PLUS balances, which enables record of up to 100 000 weighings. ALIBI guarantees continuity of data collection over a long period of time.
Data Protection
Access to valuable information is possible after logging. The administrator manages the data access configuration.
Databases
The information system is based on 5 databases, which allows several users to work with several product databases, and the registered weighing results can be subject to further analysis.
The data is registered in 5 databases:
- operators (up to 10 users),
- products (up to 1000 products),
- weighings (up to 5000 weighings),
- tares (up to 100 tares),
- ALIBI memory (up to 100 000 weighings).
Communication Interfaces
The functionality of AS R1 PLUS balances can be expanded using the following communication interfaces: USB-A, USB-B, RS 232, Ethernet, Wi-Fi® (option).
The additional USB port located at the front of the balance enables easy and fast communication with the USB flash drive, it is possible to connect additional devices via this port too.
FEATURES
INTERFACES
- Ethernet
- 2×USB
- 2×RS 232
- 4 in / 4 out
- WiFi 2,4GHz b,g,n
EXTRA FUNCTION KEYS
– header printout,
– tare editing,
– footer printout,
– product selection.
ADDITIONAL INFORMATION LINE
ERGONOMIC MECHANICAL DESIGN
- Spacious weighing chamber
- Large open-door clearance
- No rims for pane guides
- Improved aluminium base
- Tool-free disassembly of the weighing chamber
LEVELLING SYSTEM
USB AT THE BALANCE FRONT
FUNCTIONS
AUTOTEST
Diagnostic function aiming at metrological parameters determination (repeatability), the parameters are determined for the actual conditions of use. When speaking of repeatability it may be also used for weighing time optimization. Autotest is operated in an automatic mode thus operator’s time is saved.
DOSING
Weighing process for which reference mass has been determined together with tolerance for its determination. Dosing tolerance is given in [%] and it is calculated in relation to the reference value thus being a permissible deviation of this process. This solution is used for weighing powders, liquids and loose materials. Dosing function performance is often supported with bargraph – load indicator. For industrial scales it is possible to use a control systems of dosing process.
PERCENT WEIGHING
Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly exceeded.
TOTALIZING
PARTS COUNTING
Function using mass measurement for determination of measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.
PEAK HOLD
STATISTICS
Statistiscs function registers and analyses performed measurements. This supplies the user withthe following information: Max and Min standard deviation, average value, variance, range et.
CHECKWEIGHING
Checkweighing function is used for checking whether the measured sample mass is within the predefined threshold values, Low [LO] and Hgh [HI]. The thresholds are given in [g] and [kg] units. Current state of a sample being measured is signalled by means of pictorgams located on a display for laboratory balances, for industrial scales Stackligt System is used. This visual +/- inspection is in operation during segregation, control or packing process of products for which mass has bees determined with a specifaied tolerance, eg. 12860 g 961
UNDER-PAN WEIGHING
The solution enables weighing with use of a stand, which is advisable for loads of non-standard shapes and dimensions and those that generate magnetic field.
GLP PROCEDURES
Diagnostic function allowing to objectively document performed measurements. GLP procedures may be either presented in a short report form or extended one.
ANIMAL WEIGHING
Process of mass determination for a product which may unwillingly reposition within the weighing pan. Mass determination in such cases requires much longer period of time when compared to typical weighing process. It is the user who defines period of time needed for control of measured product mass. The user can thus optimize the function depending on the measured product characteristics.
DENSITY DETERMINATION
Liquid or solid body density function uses the Archimedes principle and it requires a special kit to be applied. The kit when used, replaces a weighing pan. For solids mass of the sample must be determined first in the air, next in the liquid of specified density. For liquids the so called sinker, density of which has been determined, must be measured first in the air, next in the tested liquid. The density determination process is disturbed by air bubbles which may stick either to the measured sample or the sinker. In case of laboratory equipment very light solids may be measured. For industry, solid bodies density may be determined by means of under-pan weighing. Tested load of significant size may be measured this way
DATASHEET
Maximum capacity [Max] | 120 g |
Minimum load | 10 mg |
Readability [d] | 0.1 mg |
Tare range | -120 g |
Standard repeatability [5% Max] | 0.06 mg |
Standard repeatability [Max] | 0.08 mg |
Standard minimum weight (USP) | 120 mg |
Standard minimum weight (U=1%, k=2) | 12 mg |
Permissible repeatability [5% Max] | 0.09 mg |
Permissible repeatability [Max] | 0.1 mg |
Linearity | ±0.2 mg |
Stabilization time | 2 s |
Adjustment | external |
OIML Class | – |
Display | LCD (backlit) |
Communication interface | 2×RS232, 2×USB-A (interchangeable), USB-B, Wi-Fi® (option) |
Power supply | 100 ÷ 240 V AC 50 / 60 Hz |
Operating temperature | +10 ÷ +40 °C |
Weighing pan dimensions | ø100 mm |
Packaging dimensions | 495×400×515 mm |
Net weight | 7.3 kg |
Gross weight | 9.3 kg |
ACCESSORIES
- Workstation for pipettes calibration
- Granite Antivibration tables
- Antivibration tables for laboratory balances
- Professional weighing table
- ALIBI Reader
- E2R System
- R Panel
- R.Barcode
- R-LAB
- RAD-KEY
- RADWAG Connect
- RADWAG Development Studio
- labVIEW Driver
- Density determination kit
- Under-Pan Weighing Rack
- Receipt Printer
- Antistatic ionizer
- AP2 – 1 Current Loop Unit
- RS 232 cables (scale-EPSON Printer)
- RS 232, RS 485 cables
- Barcode Scanners
- USB cable (scale – printer)
- Cigarette lighter receptacle power supply cables
- Holder for laboratory flasks
- Holder for test tubes and filters
- Displays
- Power Adapters
- RPANEL BOX
Ulasan